Model Number |
PDJ-3000A |
Specification:
ECG
Configuration: 5 lead
Lead input: 5 lead: RA, LA, RL, LL, V
Lead Selection: I,II, III, avR, avL, avF, V, MCL1
Gain: 4 grades optional
Sweep speed: 3 grades optional
Leakage of current: less than 3μA
Safety:endurance of voltage of 4000VAC/50Hz, defibrillation protection and contra electrotome interface
Indication for electrode cast: yes
Work mode: diagnosis, monitor, and surgery
Alarm indication: audible and visual, and can be reviewed
ST segment monitor: yes, and alarm range can be set.
Heart rate
Measuring range: 25 to 250 bpm
Accuracy: ±2 bmp
Tendency chart: 1~24 hours
Refurbishing rate: once for each beat
Alarm range: sustained adjustment of upper and lower limit between 20~250
Respiration
Measuring range: 1 ~ 99 bmp
Measuring error: ±2bmp
Measuring method: RA-LL Impedance
Display: waveform, respiratory rate
Alarm: yes
Alarm range: adjustable
Non-invasive pulse pressure
Measuring method: automatic vibration
Work mode: automatic/manual
Measuring parameters: SYST/DIAS/ MEAN/PR
Measuring mode: adult, pediatric, infant
Unit: mmHg or kPa
Excess pressure safety: double safe guard
Alarm range: 25 ~ 240 mmHg adjustable
SpO2
Measuring range: 0 ~100%
Accuracy: 80~100%:±2%, 60~79%:±3%, and have no requirements when less than 60%
Sensor type: double wavelength LBD
Resolution: ±1%
Alarm range: sustained adjustment of upper and lower limit between 1~100%
Alarm indication: audible and visual
Pulse rate monitor: yes
Pulse rate range: 0~255bmp
Temperature
Channel: two
Measuring range: 25~45oC
Unit: oC or F
Accuracy: ±0.2oC
Alarm range: 25~45oC
Alarm indication: yes
Performance
Display: 12color TFT display to show 4 ~ 9 waveforms
Resolution: 640×480 or 800×600
Interface: interface for CRT display and net card
Tendency observation: tendency chart of 1 - 72 hours
Alarm: upper and lower limit alarm, audible and visual, third grade alarm
Connection: Can be connected to central monitor and long-distance monitor
Memory method: manual, alarm, timing, etc.
Built-in thermal printer
|